Suggestions

DULCOTEST sensors for total chlorine

Graduated measuring ranges 0.01 – 10 mg/l

Reliably measure total chlorine with the versatile DULCOTEST sensors for total chlorine. Accurate measured values and a high degree of monitoring and process reliability are guaranteed.

Product description

Reliable online measurement of total chlorine – with DULCOTEST sensors.

Total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary refers to the total amount of free chlorine (HOCl, OCl-) and combined chlorine Mono-, di-, trichloroamine. The measuring result of type CLE (free chlorine) is subtracted from the ... more in the glossary (chloramines). Sensors for total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary are used in chloramination, chlorine elimination and wastewater monitoring. The total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary measurement is also used in conjunction with the measurement of free chlorine to measure the amount of combined chlorine Mono-, di-, trichloroamine. The measuring result of type CLE (free chlorine) is subtracted from the ... more in the glossary in swimming pool water.

Our DULCOTEST sensors for total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary are available in the versatile CTE2 series, which has different measuring ranges and signal outputs. The CTE3 series is suitable for special use in the disinfection process of inline electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary (tubular cell electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary).

Your benefits

  • More accurate and long-term stable total chlorine measurement and determination of combined chlorine thanks to the innovative design of the sensor membrane
  • Cost savings due to longer maintenance intervals when changing the membrane cap
  • Versatile use in a wide range of water qualities and in combination with many disinfection processes; CTE3 also works trouble-free together with inline electrolysis (tubular cell electrolysis); can also be used at high pH values of up to 9.5
  • Versatile use with a wide range of water qualities (pH, salinity, temperature, chemical and contamination load); in combination with many disinfection processes (chlorine gas, sodium hypochlorite, calcium hypochlorite, chloramines); can also be used at high pH values up to 9.5
  • Precise, real-time amperometric measurement for efficient process control (short response time)
  • Amperometric measuring eliminates interference caused by turbidity or colouration
  • Integrated temperature compensation eliminates interference caused by influence of temperature
  • Diaphragm-covered electrodes for reduced dependence on flow, substances in water and film-forming media
  • Diaphragm-covered measuring electrodes embedded in an electrolyte ensure long service life. This maintains optimum measuring conditions regardless of process conditions.

Field of application

The following list is a general representation of basically suitable applications for this product. Please contact your personal representative to discuss the use of the product in your individual application.

Chlorine measurement in potable water, swimming pool water, industrial water, process water and waste water

    Technical Details

    Total chlorine measuring range (different versions)

    • 0.01...20.0 mg/l
    • pH range 5.5...9.5
    • Temperature 5…45 ºC
    • Pressure max. 3.0 bar

    Sensor for Total Chlorine CTE 2-mA

    Sensor for total chlorine, including, for example, free chlorine, chloramines etc. even with high pH values in different kinds of water. For use on controllers with mA input

    Your Benefits

    • Measured variables: Total chlorine, chlorine compounds in which chlorine acts as an oxidising substance, e.g. free chlorine (HOCl and OCl-), chloramines, etc.
    • Improved measurement accuracy due to equalisation of the sensitivities for free and combined chlorine
    • Increased long-term stability due to membrane-covered sensor with innovative membrane design. This also prevents interference from changing flow rates or water constituents
    • Hydrophilic membrane ensures permeability for different water-soluble oxidising agents to the measuring electrode
    • Special reaction system of the electrolyte enables the determination of components that contain oxidising chlorine as well as use at high pH of up to 9.5

    Measured variable

    Total chlorine

    Reference method

    DPD4

    pH-range

    5.5…9.5

    Temperature

    41…113 °F (5…45 °C)

    Max. pressure

    44 psi

    Flow

    DGMa, DLG III: 7.93 ... 15.85 gph (30 ... 60 l/h)
    BAMa: 1.32 ... 26.42 gph (5 ... 100 l/h) (depending on design)

    Supply voltage

    16...24 V DC (2-wire)

    Output signal

    Uncalibrated, not temperature-compensated, not electrically isolated

    Electrical Connection

    with 2-wire signal cable via 4-pin connector on the sensor and open ends on the measuring device

    Selectivity

    Non-selective, cross-sensitive towards many oxidation agents

    Disinfection process

    Chlorine gas, hypochlorite, electrolysis with diaphragm, monochloramine. Not suitable for inline electrolysis (tubular cell electrolysis, use type CTE3 for this).

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DLG III

    Controllers

    D1C, DAC, AEGIS II, diaLog X

    Typical applications

    Drinking water, industrial water, process water, waste water. In swimming pools in combination with sensors for free chlorine to determine the combined chlorine

    Resistance to

    surfactants

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CTE 2-mA-0,5 ppm

    0.01…0.5 mg/l

    1136433

    CTE 2-mA-2 ppm

    0.02…2.0 mg/l

    1133340

    CTE 2-mA-5 ppm

    0.05…5.0 mg/l

    1136464

    CTE 2-mA-10 ppm

    0.10…10.0 mg/l

    1133338

    CTE 2-mA-20 ppm

    0.20…20.0 mg/l

    1136465

    Sensor for Total Chlorine CTE 2-CAN

    Long-term stable sensor for total chlorine, including free chlorine and chloramines. Reliable measurement even at high pH values in various waters. For operation with measuring and control devices with CAN bus connection

    Your Benefits

    • Measured variables: Total chlorine, chlorine compounds in which chlorine acts as an oxidising substance, e.g. free chlorine (HOCl and OCl-), chloramines, etc.
    • Improved measurement accuracy due to equalisation of the sensitivities for free and combined chlorine
    • Increased long-term stability due to membrane-covered sensor with innovative membrane design. This also prevents interference caused by changing flow rates or water constituents
    • Hydrophilic membrane ensures permeability for different water-soluble oxidising agents to the measuring electrode
    • Special reaction system of the electrolyte enables the determination of components containing oxidising chlorine as well as use at high pH of up to 9.5
    • Operation on the CAN bus with all the associated advantages

    Measured variable

    Total chlorine

    Reference method

    DPD4

    pH-range

    5.5…9.5

    Temperature

    41…113 °F (5…45 °C)

    Max. pressure

    44 psi

    Flow

    DGMa, DLG III: 7.93 ... 15.85 gph (30 ... 60 l/h)
    BAMa: 1.32 ... 26.42 gph (5 ... 100 l/h) (depending on design)

    Supply voltage

    Via CAN interface (11 – 30 V DC)

    Output signal

    Uncalibrated, temperature compensated, electrically isolated

    Electrical Connection

    via CAN signalling cable with M12, 5-pin plug on the sensor and M12, 5-pin socket on the measuring device

    Selectivity

    Non-selective, cross-sensitive towards many oxidation agents

    Disinfection process

    Chlorine gas, hypochlorite, electrolysis with diaphragm, monochloramine. Not suitable for inline electrolysis (tubular cell electrolysis, use type CTE3 for this).

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DLG III

    Controllers

    DULCOMARIN 3, DULCOMARIN II only with hardware after 06.02.2014 from software version 3035 or later

    Typical applications

    Drinking water, industrial water, process water, waste water. In swimming pools in combination with sensors for free chlorine to determine the combined chlorine

    Resistance to

    surfactants

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CTE 2-CAN-10 ppm

    0.01…10.0 mg/l

    1136030

    Sensor for Total Chlorine CTE 3-mA

    Sensor for total chlorine, including, for example, free chlorine, chloramines etc. even with high pH values in different kinds of water. Trouble-free use in inline electrolysis (tubular cell electrolysis). For use on controllers with mA input

    Your Benefits

    • Measured variables: Total chlorine, chlorine compounds in which chlorine acts as an oxidising substance, e.g. free chlorine (HOCl and OCl-), chloramines, etc.
    • Suitable for the determination of combined chlorine (total chlorine minus free chlorine) in the pool & wellness sector if inline electrolysis (tubular cell electrolysis) is used for chlorination
    • Improved measuring accuracy by equalising the sensitivities for free and combined chlorine
    • Low flow dependency and avoidance of interference from water constituents, as the measuring electrodes are protected by an innovative membrane system
    • Hydrophilic membrane ensures permeability for different water-soluble oxidising agents to the measuring electrode
    • Special reaction system of the electrolyte enables the determination of components that contain oxidising chlorine, as well as use at high pH of up to 9.5

    Measured variable

    Total chlorine

    Reference method

    DPD4

    pH-range

    5.5…9.5

    Temperature

    41…113 °F (5…45 °C)

    Max. pressure

    44 psi

    Flow

    DGMa, DLG III: 7.93 ... 15.85 gph (30 ... 60 l/h)
    BAMa: 1.32 ... 26.42 gph (5 ... 100 l/h) (depending on design)

    Supply voltage

    16...24 V DC (2-wire)

    Output signal

    Uncalibrated, not temperature-compensated, not electrically isolated

    Electrical Connection

    with 2-wire signal cable via 4-pin connector on the sensor and open ends on the measuring device

    Selectivity

    Non-selective, cross-sensitive towards many oxidation agents

    Disinfection process

    Chlorine gas, hypochlorite, electrolysis with membrane, inline electrolysis (tubular cell electrolysis), monochloramine

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DLG III

    Controllers

    D1C, DAC, AEGIS II, diaLog X

    Typical applications

    Determination of combined chlorine in the swimming pool in combination with sensors for free chlorine using the differential method. Also suitable for the disinfection process: Inline electrolysis (tubular cell electrolysis)

    Resistance to

    surfactants

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CTE 3-mA-2 ppm

    0.02…2.0 mg/l

    1133132

    CTE 3-mA-5 ppm

    0.05…5.0 mg/l

    1136466

    CTE 3-mA-10 ppm

    0.10…10.0 mg/l

    1133337

    CTE 3-mA-20 ppm

    0.20…20.0 mg/l

    1136467

    Sensor for Total Chlorine CTE 3-CAN

    Sensor for total chlorine, including, for example, free chlorine, chloramines etc. even with high pH values in different kinds of water. For use on controllers with mA input

    Your Benefits

    • Measured variables: Total chlorine, chlorine compounds in which chlorine acts as an oxidising substance, e.g. free chlorine (HOCl and OCl-), chloramines, etc.
    • Suitable for the determination of combined chlorine (total chlorine minus free chlorine) in the pool & wellness sector if inline electrolysis (tubular cell electrolysis) is used for chlorination
    • Improved measuring accuracy by equalising the sensitivities for free and combined chlorine
    • Low flow dependency and avoidance of interference from water constituents, as the measuring electrodes are protected by an innovative membrane system
    • Hydrophilic membrane ensures permeability for different water-soluble oxidising agents to the measuring electrode
    • Special reaction system of the electrolyte enables the determination of components containing oxidising chlorine as well as use at high pH of up to 9.5
    • Operation on the CAN bus with all the associated benefits

    Measured variable

    Total chlorine

    Reference method

    DPD4

    pH-range

    5.5…9.5

    Temperature

    41…113 °F (5…45 °C)

    Max. pressure

    44 psi

    Flow

    DGMa, DLG III: 7.93 ... 15.85 gph (30 ... 60 l/h)
    BAMa: 1.32 ... 26.42 gph (5 ... 100 l/h) (depending on design)

    Supply voltage

    Via CAN interface (11 – 30 V DC)

    Output signal

    Uncalibrated, temperature compensated, electrically isolated

    Electrical Connection

    via CAN signalling cable with M12, 5-pin plug on the sensor and M12, 5-pin socket on the measuring device

    Selectivity

    Non-selective, cross-sensitive towards many oxidation agents

    Disinfection process

    Chlorine gas, hypochlorite, electrolysis with membrane, inline electrolysis (tubular cell electrolysis), monochloramine

    Process integration

    Bypass: open sample water outlet

    Sensor fitting

    BAMa, DLG III

    Controllers

    DULCOMARIN 3, DULCOMARIN II only with hardware after 06.02.2014 from software version 3035 or later

    Typical applications

    Determination of combined chlorine in the swimming pool in combination with sensors for free chlorine using the differential method. Also suitable for the disinfection process: Inline electrolysis (tubular cell electrolysis)

    Resistance to

    surfactants

    Measuring principle, technology

    Amperometric, 2 electrodes, diaphragm-covered

    Measuring range

    Order no.

    CTE 3-CAN-10 ppm

    0.10…10.0 mg/l

    1136031

    Downloads for DULCOTEST sensors for total chlorine

    More in the Download Centre

    0 Results
    Filter by document type
    Application Example / Reference (0)
    Brochure / Flyer (0)
    Catalogue (0)
    Certificate / Declaration (0)
    Data Sheet (0)
    Drawing / Dimensional Drawing / 3D (0)
    Exploded View / Spare Parts (0)
    Material Safety Data Sheet (0)
    Operating Instructions (0)
    Others (0)
    Press Release (0)
    Publication / Article (0)
    Software (0)
    Wiring Diagram (0)

    Selected filters:

    Matching downloads

    Sort by Document type
    Load more matches ...
    No more matches available!

    Unfortunately, we did not find an appropriate match.

    All words written correctly?
    Search again with fewer or different search terms?

    Please try again later

    Service Unavailable

    The server is temporarily unable to service your request due to maintenance downtime or capacity problems. Please try again later.

    Overview

    Reliable online measurement of total chlorine – with DULCOTEST sensors.

    Total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary refers to the total amount of free chlorine (HOCl, OCl-) and combined chlorine Mono-, di-, trichloroamine. The measuring result of type CLE (free chlorine) is subtracted from the ... more in the glossary (chloramines). Sensors for total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary are used in chloramination, chlorine elimination and wastewater monitoring. The total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary measurement is also used in conjunction with the measurement of free chlorine to measure the amount of combined chlorine Mono-, di-, trichloroamine. The measuring result of type CLE (free chlorine) is subtracted from the ... more in the glossary in swimming pool water.

    Our DULCOTEST sensors for total chlorine Total of free and combined chlorine. Recommended sensor: Type CTE, reference method DPD4. more in the glossary are available in the versatile CTE2 series, which has different measuring ranges and signal outputs. The CTE3 series is suitable for special use in the disinfection process of inline electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary (tubular cell electrolysis Chlorine and sodium hydroxide are produced in-situ in electrolysis by passing an electric ... more in the glossary).

    Your benefits

    • More accurate and long-term stable total chlorine measurement and determination of combined chlorine thanks to the innovative design of the sensor membrane
    • Cost savings due to longer maintenance intervals when changing the membrane cap
    • Versatile use in a wide range of water qualities and in combination with many disinfection processes; CTE3 also works trouble-free together with inline electrolysis (tubular cell electrolysis); can also be used at high pH values of up to 9.5
    • Versatile use with a wide range of water qualities (pH, salinity, temperature, chemical and contamination load); in combination with many disinfection processes (chlorine gas, sodium hypochlorite, calcium hypochlorite, chloramines); can also be used at high pH values up to 9.5
    • Precise, real-time amperometric measurement for efficient process control (short response time)
    • Amperometric measuring eliminates interference caused by turbidity or colouration
    • Integrated temperature compensation eliminates interference caused by influence of temperature
    • Diaphragm-covered electrodes for reduced dependence on flow, substances in water and film-forming media
    • Diaphragm-covered measuring electrodes embedded in an electrolyte ensure long service life. This maintains optimum measuring conditions regardless of process conditions.

    Application

    Technical Data

    Downloads

    Contact us

    ProMinent Fluid Controls Ltd.


    »We are always available to help. Don't hesitate to contact us.«

    +1 888 709 9933 

    Matching Products

    ORP measurements with DULCOTEST sensors: Precise, reliable and application-adapted measured values in real time. Can be used in water treatment or industrial process waters with strict requirements.

    more

    The solenoid-driven diaphragm metering pump gamma incorporates a wealth of eXcellent ingenuity! With integrated pressure measurement, it ensures the smooth running of your metering process. The gamma/ X is ideal for all metering work involving liquid media.

    more

    The DULCOMETER DACb controller uses intelligent controller functions to close the control circuit between ProMinent DULCOTEST sensors and ProMinent metering pumps.

    more

    The DULCOMETER Type D1Cb is suitable for applications in drinking water, wastewater, and cooling water treatment. The single channel measuring and control unit D1Cb is equipped for all-important measured variables for basic applications in water treatment and with an error and calibration logbook.

    more

    Photometers measure nearly all disinfectants and the pH value based on the photometric principle. They are portable, compact and make safe, simple measurement possible.

    more

    The AEGIS II provides reliable control and offers the most flexible communication options to optimize efficiency and profitability for all your cooling, boiler, and wastewater or disinfection applications.